4-20 mA / 0-10 V analog loggers

Manual calibration

Efento 4–20 mA and 0-10 V loggers are factory-calibrated and provide accurate current measurements by default. However, if necessary, users may perform a manual calibration to further improve measurement precision. Calibration is carried out at four reference points (4 mA, 12 mA, 16 mA, and 20 mA for a 4–20 mA loggers and 1 V, 2 V, 6 V, 10 V for 0-10 V loggers) with a loop calibrator to supply the logger with a known, stable current / voltage.

To begin the manual calibration process, connect to the logger using the mobile application, then tap three dots in the upper-right corner → Advanced → Calibration → Manual calibration, select the channel you want to calibrate, and follow the on-screen instructions.

Calibration parameters

Measurements Number

The number of samples used to calculate the average. The highest and lowest readings are ignored.

Range: 3 to 32.

Startup Time

The time (in milliseconds) the device waits after powering the sensor board before starting measurements.

Range: 0 to 300,000 milliseconds.

Factor

A multiplier used to convert the raw ADC reading into the final Voltage (mV) or Current (mA) result.

Range: 0 to 4,294,967,295 (Note: this value is the real factor multiplied by 1,000,000).

Always on

Keeps the power supply to the sensor board on constantly.

  • Disabled (uses Startup Time).

  • Enabled (ignores Startup Time).

Last updated